Figure 10, Mounting the tubings – Metrohm 874 USB Oven Sample Processor User Manual
Page 27
■■■■■■■■■■■■■■■■■■■■■■
3 Installation
874 Oven Sample Processor
■■■■■■■■
19
1
1
2
2
3
3
4
6.1805.010
6.1805.010
6.1805.050
Figure 10
Mounting the tubings
Mount the FEP tubings as follows:
1 Load the two drying flasks that have been prepared into the holders,
see previous illustration.
2 Screw one 6.1805.050 FEP tubing (18 cm length) to the gas out-
let (left-hand side of the tower, labeled to). Screw the other tubing
end on the front drying flask to the M6 connector without a dot
mark (see right-hand arrow).
3 Screw one 6.1805.010 FEP tubing (13 cm length) to the gas inlet
(left-hand side of the tower, labeled from). Screw the other tubing
end on the rear drying flask to the M6 connector with a dot mark
(see left-hand arrow).
4 Screw the second 6.1805.010 FEP tubing (13 cm length) tightly
onto the remaining M6 connectors of the drying flasks.
The marking Drying flask on the left-hand side of the tower dis-
plays the diagram for the tubing.
Note
Tighten the screw connections well by hand.
- 915 KF Ti-Touch (382 pages)
- 800 Dosino (53 pages)
- 767 Calibrated Reference (23 pages)
- 940 Professional IC Vario ONE/SeS/Prep 2 (54 pages)
- 754 Dialysis Unit (49 pages)
- 815 Robotic Soliprep for LC (76 pages)
- Vision Manual (207 pages)
- tiamo 2.1 Manual (1532 pages)
- 825 Lab Link (37 pages)
- 808 Titrando (70 pages)
- 902 Titrando (52 pages)
- 756 KF Coulometer (163 pages)
- 756 KF Coulometer (162 pages)
- 940 Professional IC Vario ONE/LPG (98 pages)
- 850 Professional IC Anion MCS Prep 3 (154 pages)
- 850 Professional IC Anion MCS Prep 3 (152 pages)
- 904 Titrando (58 pages)
- 850 Professional IC Anion MSM-HC MCS Prep 2 (150 pages)
- 930 Compact IC Flex Oven/ChS/Deg (47 pages)
- 872 Extension Module Liquid handling (64 pages)
- 814 USB Sample Processor (90 pages)
- 814 USB Sample Processor (91 pages)
- 940 Professional IC Vario (43 pages)
- Vision – Tutorial (40 pages)
- 799 GPT Titrino (242 pages)
- 889 IC Sample Center (68 pages)
- 761 Compact IC (228 pages)
- 851 Titrando (100 pages)
- 748 DH Sample Changer (32 pages)
- 940 Professional IC Vario ONE/SeS/HPG (51 pages)
- 896 Professional Detector – Amperometry (62 pages)
- 877 Titrino plus (139 pages)
- 881 Compact IC pro – Anion (129 pages)
- 940 Professional IC Vario ONE/ChS/HPG (112 pages)
- 930 Compact IC Flex Deg (41 pages)
- 840 PC Control 5.0 / Touch Control (351 pages)
- 940 Professional IC Vario ONE/Prep 1 (45 pages)
- 776 Dosimat (42 pages)
- 717 Sample Changer (36 pages)
- 815 Robotic USB Sample Processor XL (113 pages)
- 815 Robotic USB Sample Processor XL (114 pages)
- 940 Professional IC Vario ONE/SeS/PP (126 pages)
- 838 Advanced Sample Processor Installation Instructions (109 pages)
- 700 Dosino (55 pages)
- 719 S Titrino (152 pages)