3 Installation

874 Oven Sample Processor

■■■■■■■■

19

1

1

2

2

3

3

4

6.1805.010

6.1805.010

6.1805.050

Figure 10

Mounting the tubings

Mount the FEP tubings as follows:

1 Load the two drying flasks that have been prepared into the holders,

see previous illustration.

2 Screw one 6.1805.050 FEP tubing (18 cm length) to the gas out-

let (left-hand side of the tower, labeled to). Screw the other tubing
end on the front drying flask to the M6 connector without a dot
mark (see right-hand arrow).

3 Screw one 6.1805.010 FEP tubing (13 cm length) to the gas inlet

(left-hand side of the tower, labeled from). Screw the other tubing
end on the rear drying flask to the M6 connector with a dot mark
(see left-hand arrow).

4 Screw the second 6.1805.010 FEP tubing (13 cm length) tightly

onto the remaining M6 connectors of the drying flasks.

The marking Drying flask on the left-hand side of the tower dis-
plays the diagram for the tubing.

Note