beautypg.com

Figure 2, Rear 874 oven sample processor – Metrohm 874 USB Oven Sample Processor User Manual

Page 18

background image

■■■■■■■■■■■■■■■■■■■■■■

10

■■■■■■■■

874 Oven Sample Processor

USB 2

1

2

3

4

5

6

7

8

9

Figure 2

Rear 874 Oven Sample Processor

1

USB connector (USB 1 and USB 2)
USB ports (type A) for connecting a printer,
USB hub, Titrandos, additional USB devices
etc.

2

Controller connector (Controller)
For connecting a computer with installed PC
software.

3

MSB connector (MSB 1 to MSB 3)
Metrohm Serial Bus. For connecting external
dosing devices, stirrers or remote boxes.
Mini DIN, 9-pin.

4

Mains connection socket

5

Type plate
Contains specifications concerning mains
voltage, instrument type and serial number.

6

Warning symbol
Biological hazard.(see Chapter 1.3.4, page
5)

7

Air/nitrogen connector
With M6 interior thread. Inlet for external
gassing.

8

Gas inlet and outlet
For connecting drying flasks, with M6 inner
thread.

9