5.1 Technical specifications

837 IC (Eluent/Sample/Combi) Degasser/ Instructions for Use 8.837.1003

21

5 Appendix

5.1 Technical

specifications

Provided that nothing to the contrary is mentioned, the

published data are typical values for the 837 IC (Elu-

ent/Sample/Combi) Degasser at an ambient tempera-

ture of 25°C.

Vacuum

Pressure range

< 65 mm Hg

Generation time after start

< 30 s

Capillary in the degasser

Capillary volume

Eluent:

195 µL

Sample: 670 µL

Material

Teflon AF

Solvent stability

Excellent stability against most solvents (Excep-

tion: PFCs)

Ambient temperature

Nominal operating range

+5..+45 °C (at max. 85 % atmospheric humidity)

Storage

-40...+70 °C

Transport

-40...+70 °C

Mains connection

Voltage

100…240 V (± 10%)

Frequency

50...60 Hz

Power consumption

25 W

Fuses

5 mm Ø, 20 mm long

1.0 AT (slow blow); must only be replaced by the

same type.

Safety specifications

Construction and testing

According to IEC/EN 61010-1 / UL 3101-1,

degree of protection IP30

Safety directions

This "Instructions for Use" include information and

warnings which must be heeded by the user to

assure safe operation of the instrument.