2.5 Mains connection
837 IC (Eluent/Sample/Combi) Degasser/ Instructions for Use 8.837.1003
15
2.5 Mains
connection
Please observe the following rules when connecting the instrument to
the electricity supply. If the instrument is operated with an incorrectly
set mains voltage and/or an incorrect mains fuse then it represents a
fire hazard!
2.5.1 Mains
voltage
All power supplies from 100 V to 240 V with 50/60 Hz can be connected
to the mains input of the 837 IC (Eluent/Sample/Combi) Degasser.
Adaptation to the mains supply used is carried out automatically.
2.5.2 Fuse
A 1.0 A fuse (slow blow) is contained in the fuse holder (11) of the 837
IC (Eluent/Sample/Combi) Degasser.
Make sure that the instrument is never operated with a different type of
fuse as otherwise it represents a fire hazard!
Changing the fuse
1 Pull out mains cable
Switch off the instrument and remove the mains cable from
mains supply connection (13) of the 837 IC (Eluent/Samp-
le/Combi) Degasser.
2 Remove the fuse holder
Use a screwdriver to loosen the fuse holder beside the mains
supply connection (see Fig. 4) and remove it completely.
3 Changing the fuse
Change the fuse if necessary and replace it in the fuse holder.
Use only a fuse with the following specifications:
1.0 A fuse (slow blow) Metrohm-No. U.600.0016
4 Re-insert fuse holder