3 operation, 1 general information, 2 routine operation – Metrohm 837 IC Degasser User Manual

Page 23: 1 manual operation, 2 control with «ic net, Operation, General information, Routine operation, Manual operation, Control with «ic net

background image

3.1 General information

837 IC (Eluent/Sample/Combi) Degasser/ Instructions for Use 8.837.1003

17

3 Operation

3.1 General

information

The vacuum pump works at two levels. The first level is used

to generate the vacuum, the second level to maintain it.

The display mm Hg <90 lights up orange if the pressure is

between 65 and 90 mm Hg.

The display Abs <90 lights up green, if the pressure is below

65 mm Hg. This is the normal working status, and should be

reached a few seconds after switching on the pump.

The display Status lights up green, if the pump is switched on.

A new 837 IC (Eluent/Sample/Combi) Degasser device needs a cer-

tain running-in period to reach the optimal performance. Therefore all

chambers should be run with eluent (for eluent flow paths) resp. ul-

trapure water (for sample flow paths) for one night before using it the

first time.

3.2 Routine

operation

3.2.1 Manual

operation

If you want to operate the Degasser manually, proceed as follows:

1 Installation

Install the 837 IC (Eluent/Sample/Combi) Degasser as described

in section 2.

2 Deactivate the remote-connection

Set the remote-switch 10 to OFF. The remote-connection is deac-

tivated that way.

3 Switch on/off pump

Set the mains switch 12 to I to switch on the pump and to 0 to

switch it off again.

3.2.2

Control with «IC Net»

If you want to control the degasser with «IC Net», install it as described

in section 2.3. The control via «IC Net» is described in section 6.21 of

the «IC Net 2.3» Software-Manual.