3 connection to a modular ic system, 1 electrical connection to 830 ic interface, 2 settings in «ic net – Metrohm 837 IC Degasser User Manual
Page 19: Fig. 5
2.3 Connection to a modular IC system
837 IC (Eluent/Sample/Combi) Degasser/ Instructions for Use 8.837.1003
13
2.3
Connection to a modular IC system
To use the 837 IC (Eluent/Sample/Combi) Degasser in a modular IC-
system connect it via an 830 IC Interface.
2.3.1 Electrical
connection to 830 IC Interface
Always switch off the 837 IC (Eluent/Sample/Combi) Degasser before
making new electrical connections.
The
837 IC (Eluent/Sample/Combi) Degasser is normally integrated
into the modular IC system via 830 IC Interface. Connect the remote-
interface (14) of the degasser with a free event line at the 830 IC Inter-
face using the cable 6.2128.180 according to the figure below. To acti-
vate the communication between the 830 IC Interface and the 837 IC
(Eluent/Sample/Combi) Degasser, the remote-switch (10) on the De-
gasser must be set to ON.
Fig. 5:
Connection of 837 IC (Eluent/Sample/Combi) Degasser to 830 IC
Interface
One cable 6.2128.180 is needed for each remote-connection. At the
830 IC Interface, connections can be established in two systems with 7
lines each. It doesn’t matter which event lines are used for this, you just
need to allocate them correctly when setting up the system in the soft-
ware «IC Net» (see also section 6.1 of the «IC Net 2.3» Software-
Manual).
2.3.2
Settings in «IC Net»
Add the device to the system, as described in section 6.1 of the «IC Net
2.3» Software-Manual. The settings and control elements for the Degas-
ser in the «IC Net» are described in section 6.21 of the «IC Net 2.3» Soft-
ware-Manual.