beautypg.com

3 connection to a modular ic system, 1 electrical connection to 830 ic interface, 2 settings in «ic net – Metrohm 837 IC Degasser User Manual

Page 19: Fig. 5

background image

2.3 Connection to a modular IC system

837 IC (Eluent/Sample/Combi) Degasser/ Instructions for Use 8.837.1003

13

2.3

Connection to a modular IC system

To use the 837 IC (Eluent/Sample/Combi) Degasser in a modular IC-

system connect it via an 830 IC Interface.

2.3.1 Electrical

connection to 830 IC Interface

Always switch off the 837 IC (Eluent/Sample/Combi) Degasser before

making new electrical connections.

The

837 IC (Eluent/Sample/Combi) Degasser is normally integrated

into the modular IC system via 830 IC Interface. Connect the remote-

interface (14) of the degasser with a free event line at the 830 IC Inter-

face using the cable 6.2128.180 according to the figure below. To acti-

vate the communication between the 830 IC Interface and the 837 IC

(Eluent/Sample/Combi) Degasser, the remote-switch (10) on the De-

gasser must be set to ON.

Fig. 5:

Connection of 837 IC (Eluent/Sample/Combi) Degasser to 830 IC
Interface

One cable 6.2128.180 is needed for each remote-connection. At the

830 IC Interface, connections can be established in two systems with 7

lines each. It doesn’t matter which event lines are used for this, you just

need to allocate them correctly when setting up the system in the soft-

ware «IC Net» (see also section 6.1 of the «IC Net 2.3» Software-

Manual).

2.3.2

Settings in «IC Net»

Add the device to the system, as described in section 6.1 of the «IC Net

2.3» Software-Manual. The settings and control elements for the Degas-

ser in the «IC Net» are described in section 6.21 of the «IC Net 2.3» Soft-

ware-Manual.