5.11 Metrohm Suppressor Module (MSM)

■■■■■■■■■■■■■■■■■■■■■■

80

■■■■■■■■

882 Compact IC plus – Anion – MCS

1

2

3

Figure 37

Pump tubing connection – Changing the filter

1

Tubing olive

2

Filter 6.2821.130
Packaging contains 10 items.

3

Filter housing

Replacing the filter

1 Unscrewing filter screw

Screw the filter screw (37-3) out of the tubing olive (37-1) with
the aid of two 6.2621.000 adjustable wrenches.

2 Replacing the filter

Remove the old filter (37-2) with tweezers.

Place the new filter (37-2) flat in the tubing olive (37-1) with
tweezers.

3 Mounting filter screw

Screw the filter screw (37-3) back into the tubing olive (37-1) and
tighten by hand. Then additionally tighten with two 6.2621.000
adjustable wrenches.

5.11

Metrohm Suppressor Module (MSM)

5.11.1

Protection

To protect the suppressor against foreign particles or bacterial growth, a
pump tubing connection with filter (6.2744.180) (see Figure 22, page 42)
must be mounted between the peristaltic pump (see Figure 20, page 40)
and the inlet capillaries of the suppressor.