11 metrohm suppressor module (msm), 1 protection, Protection – Metrohm 882 Compact IC plus – Anion – MCS User Manual
Page 90: Figure 37, Pump tubing connection – changing the filter

5.11 Metrohm Suppressor Module (MSM)
■■■■■■■■■■■■■■■■■■■■■■
80
■■■■■■■■
882 Compact IC plus – Anion – MCS
1
2
3
Figure 37
Pump tubing connection – Changing the filter
1
Tubing olive
2
Filter 6.2821.130
Packaging contains 10 items.
3
Filter housing
Replacing the filter
1 Unscrewing filter screw
■
Screw the filter screw (37-3) out of the tubing olive (37-1) with
the aid of two 6.2621.000 adjustable wrenches.
2 Replacing the filter
■
Remove the old filter (37-2) with tweezers.
■
Place the new filter (37-2) flat in the tubing olive (37-1) with
tweezers.
3 Mounting filter screw
■
Screw the filter screw (37-3) back into the tubing olive (37-1) and
tighten by hand. Then additionally tighten with two 6.2621.000
adjustable wrenches.
5.11
Metrohm Suppressor Module (MSM)
5.11.1
Protection
To protect the suppressor against foreign particles or bacterial growth, a
pump tubing connection with filter (6.2744.180) (see Figure 22, page 42)
must be mounted between the peristaltic pump (see Figure 20, page 40)
and the inlet capillaries of the suppressor.