Metrohm 882 Compact IC plus – Anion – MCS User Manual
Page 23

■■■■■■■■■■■■■■■■■■■■■■
3 Installation
882 Compact IC plus – Anion – MCS
■■■■■■■■
13
8 Connecting the instrument
■
Connect the instrument to a computer on which the software
MagIC Net™ is installed using the USB cable (6.2151.020) (see
Chapter 3.16.1, page 52).
■
Connect the instrument to the mains supply (see Chapter 3.16.2,
page 52).
9 Initial start-up
■
Switch on the PC and start the software MagIC Net™.
■
Switch on the instrument.
■
Deaerate the high pressure pump (see Chapter 3.9.2, page 32).
■
Set contact pressure of the peristaltic pump (see "Set flow rate",
page 44).
■
Rinse the instrument without column with eluent for 5 minutes.
10 Installing guard and separation column
■
Remove the coupling (6.2744.040) between the column inlet
capillary and the capillary of the MSM labeled with in.
■
Optional Connect guard column (see Chapter 3.17, page 53)
– Fasten the guard column to the end of the column inlet
capillary (see leaflet to the guard column).
– Rinse the guard column with eluent for approx. 5 minutes.
■
Connect the separation column (see Chapter 3.18, page 54)
– Connect the inlet oft the separation column either with the
end of the column input capillary using a short pressures
screw (6.2744.070).
OR
Connect the inlet of the separation column with the guard
column (if used) (see leaflets to the separation column and
the guard column)
– Connect the MSM capillary labeled with in with the output
of the separation column using a short pressure screw
(6.2744.070).
■
Hang separation column with chip in the column holder of the
instrument.
11 Conditioning the instrument
The instrument is now ready for measuring samples.