3.12 Injection valve
■■■■■■■■■■■■■■■■■■■■■■
38
■■■■■■■■
882 Compact IC plus – Anion – MCS
2 Mounting new sample loop
■
Fasten one end of the sample loop (18-2) with a 6.2744.010
PEEK pressure screw (18-7) to connector 3.
■
Fasten the other end of the sample loop (18-2) with a second
6.2744.010 PEEK pressure screw (18-7) to connector 6.
3.12.2
Mode of operation of the injection valve
The injection valve (see Figure 19, page 38) can adopt two valve posi-
tions - FILL and INJECT. Switching back and forth between the two valve
positions determines whether the sample path or the eluent path is gui-
ded through the sample loop. The following figure provides a schematic
display of the flow paths of the two valve positions.
1
2
4
5
3
6
1
2
4
5
3
6
1
2
4
3
5
4
3
1
2
A
B
Figure 19
Injection valve – Positions
A
Position FILL
B
Position INJECT
1
Eluent input
Capillary coming from the high pressure
pump.
2
Eluent output
Capillary to the column.
3
Sample input
Sample aspirating capillary.
4
Sample output
Capillary to waste container.
5
Sample loop
Position A
In the position FILL, the sample solution flows
through the sample loop to the waste container.
The eluent flows directly to the separation col-
umn at the same time.