10 metrohm co2 suppressor (mcs), 11 mains connection, 12 interfaces – Metrohm 882 Compact IC plus – Anion – MCS User Manual
Page 108: 10 metrohm co, Suppressor (mcs)

7.10 Metrohm CO
2
Suppressor (MCS)
■■■■■■■■■■■■■■■■■■■■■■
98
■■■■■■■■
882 Compact IC plus – Anion – MCS
7.10
Metrohm CO
2
Suppressor (MCS)
Material
fluoropolymer
Resistance to sol-
vents
No restriction (apart from PFC)
Vacuum
Working area
Microprocessor-controlled / stabilized
Build-up time
after start
< 30 s
Capillary volume
400 µL
Recommended
flow range
0.1…1.0 mL
7.11
Mains connection
Required voltage
100…240 V ± 10 % (auto-sensing)
Required fre-
quency
50…60 Hz ± 3 (auto-sensing)
Power consump-
tion
■
65 W for typical analysis application
■
25 W standby (conductivity detector to 40 °C)
Power supply unit
■
Up to 300 W maximum, electronically monitored
■
internal fuse 3.15 A
7.12
Interfaces
USB
Input
1 USB upstream, type B (for connection to the PC)
Output
2 USB downstream, type A
MSB
2 MSB MiniDin 8-pin (female) (for Dosino, stirrer, remote lines, …)
Caution
When connecting an instrument to the MSB connector you must
switch off the 882 Compact IC plus.
Detector
1 DSUB-15-pin high density (female)
Column recogni-
tion
for an intelligent column