beautypg.com

4 shutting down and starting back up, Shutting down and starting back up – Metrohm 930 Compact IC Flex Oven/SeS/PP User Manual

Page 67

background image

■■■■■■■■■■■■■■■■■■■■■■

5 Operation and maintenance

930 Compact IC Flex Oven/SeS/PP (2.930.2500)

■■■■■■■■

59

5.1.4

Shutting down and starting back up

If the instrument is not used for a long period, the whole IC system (with-
out separation column) must be rinsed to be free of salts using methanol/
ultrapure water (1:4). This prevents eluent salts from crystallizing and then
causing damage.

Rinsing the IC system to be free of salts

1 In the software, stop the hardware and wait until the pressure in the

high-pressure pump has dissipated.

2 Remove the guard column and the separation column from the elu-

ent path. Connect the connection capillaries directly with each other
using a coupling (6.2744.040).

3 Rinse the IC system for 15 minutes with methanol/ultrapure water

(1:4).

4 Optional: Only if the IC system is equipped with a ChS plug-in or a

SeS plug-in.

In the software, switch the suppressor twice during the rinsing proc-
ess (STEP command).

5 Optional: Only if the IC system is equipped with a ChS/PP plug-in or a

SeS/PP plug-in.

Rinse the two pieces of pump tubing for the peristaltic pump with
water for 5 minutes at level 3. Finish by releasing the peristaltic
pump's contact pressure.

Putting the IC system back into operation

1 Check that a coupling (6.2744.040) is installed in place of the guard

column and the separation column.

2 Rinse the IC system with eluent for 15 minutes.

3 Remove the coupling and install the guard column and the separa-

tion column and (see Chapter 3.19, page 52).