6.1 Problems and their solutions
■■■■■■■■■■■■■■■■■■■■■■
76
■■■■■■■■
850 Professional IC Cation – HP-Gradient
Problem
Cause
Remedy
Guard column – diminished
performance.
■
Replace guard column (see Chapter 3.17,
page 47).
Separation column – dead
volume at column head.
■
Install separation column in reverse flow
direction (if permitted by the information
sheet).
■
Replace separation column (see "Connect-
ing and rinsing the separation column",
page 49).
Precision problems -
significant scatter-
ing of the measured
values
Sample – Gas bubbles in
the sample.
Use sample degasser (see Chapter 3.13, page
39).
Injection valve – sample
loop.
Check installation of the sample loop (see
Chapter 3.14.1, page 41).
Sample – Rinsing volume
too low.
Prolong the rinsing time (see Chapter 5.8,
page 69).
Injection valve – defective.
Request Metrohm Service.
Unexpected change
to the retention
times in the chroma-
tograms
Separation column –
diminished separating effi-
ciency.
■
Regenerate separation column (see Chap-
ter 5.11.4, page 72).
■
Replace separation column (see "Connect-
ing and rinsing the separation column",
page 49).
Eluent – Gas bubbles in the
eluent.
Check connections of the eluent degasser (see
Chapter 3.9, page 31).
High pressure pump –
defective.
Request Metrohm Service.
Vacuum is not being
built
Eluent Degasser – Connec-
tor Vacuum on the rear of
the instrument not (tightly)
sealed.
■
Seal the connector Vacuum tightly with a
6.144.6040 threaded stopper.
or when using an 872 Extension Module:
■
Connect an FEP tubing between the IC
instrument and the Extension Module and
tighten both clamping screws until they
seal.