Metrohm 850 Professional IC Cation - Gradient User Manual
Page 21

■■■■■■■■■■■■■■■■■■■■■■
3 Installation
850 Professional IC Cation – HP-Gradient
■■■■■■■■
13
3 Connecting the eluent path
■
Equip an eluent aspiration tubing for each of the two high pres-
sure pumps and connect with the eluent bottles (see Chapter
3.8.1, page 27).
■
Instead of the column, connect the 6.2744.040 coupling with a
6.2744.070 short pressure screw at the end of the pre-installed
column inlet capillary.
■
Connect the detector inlet capillary with a 6.2744.070 short pres-
sure screw to the .6.2744.040 coupling.i
4 Connecting the sample path
Note
The sample degasser does not have to be connected. We recom-
mend the usage of the sample degasser only if the sample matrix
requires it.
■
Guide the 6.1803.040 sample aspiration capillary connected to
the sample input of the injection valve out of the instrument
through a capillary feed-through and connect it with the Sample
Processor or
■
If the sample degasser is used connect the sample aspiration
capillary with a 6.2744.090 long pressure screw to the input of
the sample degasser (see Chapter 3.13, page 39).
Connect a part of the 6.1803.040 sample aspirating capillary with
a 6.2744.090 long pressure screw to the input of the sample
degasser. Guide the other end out of the instrument through a
capillary feed-through and connect it to the Sample Processor.
5 Connecting the instrument
■
Connect the instrument to the PC with a 6.2151.020 USB cable
(see Chapter 3.16.1, page 46).
■
Connect instrument to mains supply (see Chapter 3.16, page
46).
6 Initial start-up
■
Switch on the PC and start MagIC Net.
■
Switch on the instrument.
■
Deaerate the high pressure pump (see Chapter 3.10.2, page
35).