Metrohm 850 Professional IC Cation - Gradient User Manual
Page 83

■■■■■■■■■■■■■■■■■■■■■■
6 Troubleshooting
850 Professional IC Cation – HP-Gradient
■■■■■■■■
75
Problem
Cause
Remedy
Very noisy baseline
High pressure pump – con-
taminated pump valves.
Clean pump valves (see Chapter 5.5.2, page
56).
Eluent – Leak in the eluent
path.
Check the eluent path
Eluent – Blockage in the
eluent path.
Check the eluent path
High pressure pump –
defective piston seals.
Replace piston seals (see Chapter 5.5.2, page
56).
Pulsation damper not con-
nected.
Connect the pulsation damper (see Chapter
3.12, Page 38).
Pulsation damper not con-
nected.
Connect pulsation damper (see Chapter 3.12,
page 38).
Background conduc-
tivity too high
Incorrect eluent.
Change eluent (see Chapter 5.4.2.3, page 56).
Data of the separa-
tion column cannot
be read.
Column chip contamina-
ted.
Clean contact surfaces of the column chip
(with alcohol).
Column chip defective.
1. Save column configuration in MagIC Net.
2. Notify Metrohm Service.
Individual peaks
greater than
expected
Sample – Carry-over of the
samples from previous
measurement.
Rinse system longer between two samples.
Poor reproducibility
of the retention
times
Eluent – Leak in the eluent
path.
Check the eluent path
Eluent – Blockage in the
eluent path.
Check the eluent path
Chromatograms
have poor resolution
Separation column –
diminished separating effi-
ciency.
■
Regenerate separation column (see Chap-
ter 5.11.4, page 72).
■
Replace separation column (see "Connect-
ing and rinsing the separation column",
page 49).
Extreme spread of
the peaks in the
chromatogram.
Splitting (dual
peaks)
Connections – dead vol-
ume in the system.
Check connections (see Chapter 3.5, page 16)
(use PEEK capillaries with an internal diameter
of 0.25 mm between the injection valve and
detector).