3.13 Sample degasser

■■■■■■■■■■■■■■■■■■■■■■

40

■■■■■■■■

850 Professional IC Cation – HP-Gradient

1

2

3

4

Figure 22

Sample degasser

1

Sample degasser input

2

Sample degasser output

3

PEEK pressure screw, long 6.2744.090

4

Connection capillaries 6.1803.040

Connecting the sample degasser

1 Remove and keep 6.2744.220 thread stoppers from the input and

output of the sample degasser.

2 Connect the end of the 6.1803.040 sample aspirating capillary con-

nected to the injection valve with a long PEEK pressure screw (22-3)
to the output of the sample degasser (22-2).

3 Connect 6.1803.040 connection capillary with a long PEEK pressure

screw (22-3) to the input of the sample degasser (22-1).

4 Guide the other end of the connection capillary out of the instrument

through a capillary feed-through.

Caution

If the sample degasser is not used, the input and output must be
sealed with the 6.2744.220 thread stoppers.