Metrohm 768 KF Oven User Manual
Page 28

3.2 Remote control commands
768 KF Oven
25
3.2.2.17
Info.Report
$G
Output of the selected report.
3.2.2.18
Info.Report.Select
configuration, parameters, result
Selection of the report.
configuration: Configuration report. Not accessible when a determination is running.
parameters: Parameter report. Only "live" parameters when a determination is running.
result:
Result report of the last completed determination.
The report output must be triggered with &Info.Report $G, see 3.2.2.17.
Reports which are sent automatically by the KF Oven start with a space (ASCII 32) and '.
This is followed by the individual identification for each report. Reports which are requested
via RS232 ($G) have the same individual identification of the report blocks, but are started
with ' only (no preceding space).
3.2.2.19
Info.Results.PurgeTime
read only
Info.Results.CondTime
read only
Info.Results.SampleHeatTime read only
Info.Results.LowTemp
read only
Info.Results.HighTemp
read only
Info.Results.GasFlow
read only
Info.Results.LowFlow
read only
Info.Results.HighFlow
read only
Inquiry of the current results.
.PurgeTime: Purge time.
.CondTime:
Conditioning time.
.SampleHeatTime: Time during which the sample has been heated.
.LowTemp:
Lowest temperature during the sample heating.
.HighTemp:
Highest temperature during the sample heating.
.GasFlow:
Mean gas flow during the sample heating.
.LowFlow:
Lowest gas flow during the sample heating.
.HighFlow:
Highest gas flow during the sample heating.
3.2.2.20
Info.ActualInfo.Inputs.Status
read only
Info.ActualInfo.Inputs.Change read only
Info.ActualInfo.Inputs.Clear
$G
Info.ActualInfo.Outputs.Statusread only
Info.ActualInfo.Outputs.Changeread only
Info.ActualInfo.Outputs.Clear
$G
Status sends the current status of the I/O lines. Change sends information regarding
whether a change in the status of a line has taken place since it was last cleared, Clear
clears the byte of Change.
For the output, the byte is converted to binary, e.g.
|0|0|0|0|1|0|1|0|
Line No.
|7|6|5|4|3|2|1|0|
Output: 21 + 23 = "10"
1 means ON or change; 0 means OFF or no change.
The lines are assigned as follows (see also pages 53, 54):
Inputs:
Outputs:
0 Start (Pin 21)
0 Ready (Pin 5)
1 Stop (Pin 9)
1 Start (Pin 18)
2 Terminate (Pin 22) 2 Stop (Pin 4)
3 Pin 10
3 HeatSmpl (Pin 17)
4 Pin 23
4 Terminate (Pin 3)
5 Pin 11
5 Error (Pin 16)
6 Pin 24
6 Pin 1
7 Cond.ok (Pin 12)
7 Pin 2