2 description of the remote control commands – Metrohm 768 KF Oven User Manual
Page 26

3.2 Remote control commands
768 KF Oven
23
3.2.2 Description of the remote control commands
3.2.2.1
Mode
$G, $S
Start and stop ($G, $S) of the automatic determination.
3.2.2.2
Mode.Temp
50...300
Entry of the sample temperature in °C.
3.2.2.3
Mode.Gas.UnitFlow
mL/min, L/h
Mode.Gas.MinFlow
0...5...999
.UnitFlow:
Selection of the unit for display of the gas flow.
.MinFlow:
Setting of the minimum gas flow in the unit selected above. If the flow is less
than the minimum gas flow, error E163 appears (gas flow too low).
3.2.2.4
Mode.Gas.Type.Select
air, N2, other
Mode.Gas.Type.OtherFac
0.001...1...9999
.Select:
Selection of the gas.
.Factor:
If "other" has been selected above, setting of the factor for measurement of
the gas flow, factors see page 11.
3.2.2.5
Mode.Gas.PurgeTime
0...99999
Mode.Gas.CondTime
0...99999
.PurgeTime: Purge time with valve to "Purge". Entry in s.
.CondTime:
Conditioning time with valve to "Transfer". Entry in s.
3.2.2.6
Config.OvenSet.AutoPrep
ON, OFF
ON means: Automatic heating and preparation of the oven after switching on.
3.2.2.7
Config.OvenSet.ValveControl
ON, OFF
Config.OvenSet.StartCond
ON, OFF
Config.OvenSet.TempLimit
1...5...100
Config.OvenSet.TempCorr
0...
±99.9
• ValveControl:
ON means automatic switching of the valve from PURGE to TRANSFER at the end of the
determination.
• StartCond:
ON means: Insert sample boat during automatic determination only if the connected
device is conditioned (input line pin 12 of the socket "Remote" active).
• TempLimit:
Starting a determination is possible only if the current temperature is in the range of
±X
°C from the sample temperature.
• TempCorr:
Correction of the sample temperature according to individual sample and Oven
requirements.