3.1 General rules
768 KF Oven
19
RS send errors:
E40
DSR=OFF. No proper handshake for more than 1 s.
Exit:
E41
DCD=ON. No proper handshake for more than 1 s.
Exit:
E42
CTS=OFF. No proper handshake for more than 1 s.
Exit:
E43
The transmission has been interrupted with XOFF for at least 3
s.
Exit: Send XON or
E44
The RS parameters are no longer the same for both devices.
Reset.
E45
The receive buffer contains an incomplete command (LF
missing). Sending from the instrument is therefore blocked.
Exit: Send LF or
Device-specific errors:
E26
Manual stop.
Exit: The error message disappears on the next start.
E135
Check temperature sensor of sample temperature.
Exit: Rectify fault.
E154
Sample temperature not OK.
Exit: The error message disappears when the fault is rectified
or &M $S.
E163
Gas flow too low.
Exit: The error message disappears when the fault is rectified
or &M $S.
E164
The attached titrator is not conditioned.
Exit: The error message disappears when the fault is rectified
or &M $S.
E165
The oven temperature is higher than 360 °C.
Exit: The error message disappears when the fault is rectified.
E168
The temperature sensor for the oven temperature is not OK.
Exit: Rectify fault.
E169
The gas flow sensor is not OK (flow >500 mL/min).
Exit: Rectify fault. Continuation is possible only when the
minimum gas flow is set to 0 mL/min.