3 status and error messages – Metrohm 768 KF Oven User Manual
Page 21

3.1 General rules3.1 General rules
768 KF Oven
18
3.1.3 Status and error messages
Detailed status conditions
Status conditions of the global $G:
$G.Mode .Inac
Waiting during start delay
.PurgeTime
Waiting during purge time
.CondTime
Waiting during cond.time
.HeatSmpl
Heating the sample
.Terminate
Carrying out the terminating steps
$G.Assembly .Prep.Wait
Heating the oven to the set temperature
.Boat
The sample boat has been manually moved
Status conditions of the global $R:
$R.Mode .Ready
Inactive: Ready to start an automatic determination
$R.Assembly.Ready
An Assembly step has been carried out. Ready to carry out
another Assembly step (&Mode $G will trigger error message
E31).
Status conditions of the global $S:
The instrument gives the status from which it has been stopped. The detailed status information is
therefore identical as for the global status $G.
Error messages
RS receive errors:
E36
Parity error
Exit:
E37
Stop Bit
Exit:
E38
Overrun error. At least 1 character could not be read.
Exit:
E39
The internal receive buffer is full (>82 characters).
Exit: