Metrohm TitrIC 1 User Manual

Page 14

background image

Application Bulletin 285 e
Installation Instructions of TitrIC 1
Page 14/16

6.2.6 Electrodes

pH electrode: Under

Configuration/Sensors

click on

Edit / New / pH electrode

and enter under

Sensor

name Aquatrode Plus

, type in the desired sensor information (e.g. serial number) and click on

.

Ca-ISE: Under

Configuration/Sensors

double click on the entry

ISE electrode

and type in the desired

sensor information (e.g. serial number) and click on

.

Conductivity sensor: Under

Configuration / Sensors

click on

Edit / New / other sensor

and enter under

Sensor name Conductivity sensor

type in the desired sensor information and click on

.

6.2.7 Methods

The following seven methods are provided:

TitrIC 1 - complete run:

With this method all the parameters possible with the standard instruments

are measured and the IC measurement is initiated.

TitrIC 1 - titration and measurement:

Only titrations and direct measurements are carried out; the IC

measurement is not initiated. In TitrIC under

Settings / IC Net settings

the

IC Anions

and

IC Cations

have to be deactivated, or the method is started up only from tiamo.

TitrIC 1 - preparation of titration equipment:

This method is used for preparing all titration instruments,

i.e., rinsing the tubing connections of the Dosing Units with reagents or deionized water and
rinsing the external cell. Start this method only from tiamo – not from TitrIC!

TitrIC 1 - IC calibration:

This method is used for calibrating the IC measurement with standard

solutions without having to carry out the titration part of a typical TitrIC measurement; this
means that a lot of solution can be saved.

TitrIC 1 – semi automated pH calibration:

With this method the pH electrode is calibrated automated;

the buffer solutions are poored into the titration vessel. Start this method only from tiamo – not
from TitrIC!

TitrIC 1 - pH calibration:

Here the Aquatrode is taken out of the titration vessel and placed into the

buffer solutions. Start this method only from tiamo!

TitrIC 1 - titer determination of EDTA:

This method is used for the titer determination. Start this

method only from tiamo!

TitrIC 1 - titer determination of HCl:

This method is used for the titer determination. Start this method

only from tiamo!

An automated conductivity cell calibration method for tiamo can be delivered if desired

([email protected]); it is recommend to do it manually.



Import:
Click on

Method

at the left, then select under

File / Method manager… / Edit / Import…

in the

directory

C:\Program Files\Metrohm\tiamo\methods_titric

all methods and click on

.

Note!
Alterations of these methods should only be carried out by a person at the Administrator level
and who is thoroughly familiar with tiamo. Never forget that in TitrIC this new method must be
selected in the sample table under

Methods tiamo

.