Metrohm TitrIC 1 User Manual
Page 14
Application Bulletin 285 e
Installation Instructions of TitrIC 1
Page 14/16
6.2.6 Electrodes
pH electrode: Under
Configuration/Sensors
click on
Edit / New / pH electrode
and enter under
Sensor
name Aquatrode Plus
, type in the desired sensor information (e.g. serial number) and click on
.
Ca-ISE: Under
Configuration/Sensors
double click on the entry
ISE electrode
and type in the desired
sensor information (e.g. serial number) and click on
.
Conductivity sensor: Under
Configuration / Sensors
click on
Edit / New / other sensor
and enter under
Sensor name Conductivity sensor
type in the desired sensor information and click on
.
6.2.7 Methods
The following seven methods are provided:
TitrIC 1 - complete run:
With this method all the parameters possible with the standard instruments
are measured and the IC measurement is initiated.
TitrIC 1 - titration and measurement:
Only titrations and direct measurements are carried out; the IC
measurement is not initiated. In TitrIC under
Settings / IC Net settings
the
IC Anions
and
IC Cations
have to be deactivated, or the method is started up only from tiamo.
TitrIC 1 - preparation of titration equipment:
This method is used for preparing all titration instruments,
i.e., rinsing the tubing connections of the Dosing Units with reagents or deionized water and
rinsing the external cell. Start this method only from tiamo – not from TitrIC!
TitrIC 1 - IC calibration:
This method is used for calibrating the IC measurement with standard
solutions without having to carry out the titration part of a typical TitrIC measurement; this
means that a lot of solution can be saved.
TitrIC 1 – semi automated pH calibration:
With this method the pH electrode is calibrated automated;
the buffer solutions are poored into the titration vessel. Start this method only from tiamo – not
from TitrIC!
TitrIC 1 - pH calibration:
Here the Aquatrode is taken out of the titration vessel and placed into the
buffer solutions. Start this method only from tiamo!
TitrIC 1 - titer determination of EDTA:
This method is used for the titer determination. Start this
method only from tiamo!
TitrIC 1 - titer determination of HCl:
This method is used for the titer determination. Start this method
only from tiamo!
An automated conductivity cell calibration method for tiamo can be delivered if desired
([email protected]); it is recommend to do it manually.
Import: Click on
Method
at the left, then select under
File / Method manager… / Edit / Import…
in the
directory
C:\Program Files\Metrohm\tiamo\methods_titric
all methods and click on
.
Note!
Alterations of these methods should only be carried out by a person at the Administrator level
and who is thoroughly familiar with tiamo. Never forget that in TitrIC this new method must be
selected in the sample table under
Methods tiamo
.