6 software – Metrohm TitrIC 1 User Manual
Page 12
Application Bulletin 285 e
Installation Instructions of TitrIC 1
Page 12/16
6 Software
6.1 Installation of the software
Edgeport driver – installation (Inside Out NETWORKS-CD): Connect the Edgeport with the
USB cable (included in 2.145.0320) with the PC and then insert the CD. It’s not necessary to
start windows update, click on
automatic installation of the software
.
tiamo 1.1 – installation (tiamo-CD): All the standard directories proposed by the program
should be accepted.
IC Net 2.3 – installation (Metrodata-CD): All the standard directories proposed by the program
should be accepted. Restart windows.
TitrIC 1.0 – installation (Metrodata-CD): TitrIC 1.0 must only be installed after tiamo 1.1 and
IC Net 2.3 have already been installed. After installation the standard TitrIC directory can be
found under
C:\Programme\ Metrohm\TitrIC
, the TitrIC.exe file under
C:\Programme\Metrohm\tiamo\bin
.
All programs must be shut down!
6.2 Configuration
of
tiamo
The Controller cable (6.2151.000), which is already connected to the 809, is now connected to an
USB interface on the PC. Wait a few seconds. The driver installation for 809 and 815 follows. It’s not
necessary to start windows update, click on
automatic installation of the software
for both.
Switch on the 712 Conductometer
tiamo 1.1 is started
6.2.1 Devices
Connected devices are automatically recognized when tiamo is started. After confirmation of the auto-
matically generated request the devices are stored in the configuration, keep the proposed device
names. The 712 Conductometer is not recognized automatically and must therefore be added with the
name
712_1
via
Configuration / Devices / Edit / New / Meter / 712 Conductometer
, as well as selected
COM 5
under
the entry
RS 232
and confirmed with
. Depending on the PC, the Edgeport port 1 could be assigned
to a different COM port. This is checked by clicking on
Start / System control / System / Hardware / Device
manager / Connections (COM and LPT)
and checking which COM port has been assigned to port 1 of the
Edgeport.
6.2.2 Swing Head configuration
Under
Configuration
at the top left, double-click on
815_1
under
Devices
, click on the entry
Tower 1
, set the
Axial distance
to 196.0 mm and in the middle of the window under
Swing Head
click on
Configuration
and
Yes
and then enter the following values:
Configuration of the Swing Head for Titration head 6.1462.050
Confirm the entries twice with <
OK>
and then mark
815_1
under
Devices
and click on
Edit /
Initialize
. Wait a
few seconds. The titration head (6.1462.070) can now be mounted on the Swing Head.