Application Bulletin 285 e

Installation Instructions of TitrIC 1

861 Single Channel Advanced Compact IC with 815
Robotic Sample Processor

VENT

20

mL

800 Dosino

Metrohm

VENT

10

mL

800 Dosino

Metrohm

VENT

50

mL

800 Dosino

Metrohm

6.2055.100

809

Titrando

VENT

20

mL

800 Dosino

Metrohm

712

815

The TitrIC 1 System with one 861 Compact IC, an 815 Robotic Sample Processor and 809 Titrando

The TitrIC 1 System is used for the fully automatic analysis of water samples using direct measure-
ment, titration and ion chromatography. The following measured values are determined in a very short
time: temperature, conductivity, pH, acid capacity, calcium, magnesium, total hardness and, in
parallel, the concentrations of the individual anions. Further Metrohm instruments can be incorporated
in the existing system at any time and used to measure additional parameters.

Analytical sequence

1. Conductivity is measured in the sample vessel on the sample changer.
2. First sample transfer to the 861 Advanced Compact IC.
3. The IC measurement is started automatically.
4. Second sample transfer to the external cell of the 815 Robotic Sample Processor.
5. Temperature, pH and acid capacity (p and m values) calcium, magnesium and total hardness

are determined.

The whole procedure is controlled by the TitrIC software: The user enters the sample position and
sample identification, IC system and tiamo method, after which TitrIC transfers all the relevant
information to tiamo and IC Net. When the measurement has been concluded, TitrIC automatically
generates a joint report containing all results. This report can be printed out or produced as an Excel
table.