Application Bulletin 285 e
Installation Instructions of TitrIC 1
Page 11/16

5 Miscellaneous

5.1 Stirrer

The 802 Rod Stirrer is inserted through opening 9 of the external cell (see page 7) and connected to the
tower of the 815. The propeller (6.1909.020) is pushed on from below. Connect the stirrer to the
connector at the left of membrane pump 2 of the 815.

5.2 Length of the aspiration tips in the external cell

Clamp the titration vessel (6.1415.220) to the titration vessel lid (6.1414.060).

The M8 aspiration tip in opening 1 must reach right to the base of the titration vessel as this is

the only way to ensure that the external cell is cleaned properly.

The buret tips of the 3-way stopper in opening 6 and the buret tip of opening 3 should reach the

middle of the titration vessel. The three tubings of the 3-way stopper have to be mounted with
Antidiffusion micro valves to ensure a precise titration.

5.3 Electrodes

The pH electrode (Aquatrode Plus with Pt 1000; 6.0257.000) is inserted into opening 4 together

with the SGJ sleeve (6.1236.050). The cable is connected to the 809 Titrando at

Input 2

. The

plug of the gray cable is connected to the opening marked

Ind.

. The red plugs with the adapters

6.2103.140 and 6.2103.130 are inserted in the openings marked

Temp.

If the electrode is not to be used for some time then it should be returned to its holder and kept
in the storage solution (6.2323.000).

The Ca-ISE (Polymer membrane electrode Ca, 6.0508.110) is connected to the electrode cable

(6.2104.020) and inserted into opening 8. The cable is connected to the 809 Titrando at

Input 1

.

The plug of the gray cable is connected to the opening marked

Ind.

. The red plugs with the

adapters 6.2103.140 and 6.2103.130 are inserted in the openings marked

Temp.

.

To have a reference electrode for the Ca-ISE one can short-circuit the reference channel of the
Ca-ISE with the one of the Auquatrode. Thereto connect

Ref.

of

Input 1

with

Ref.

of

Input 2

with the

electrode cable 2x plug B (6.2106.020).

Conductivity cell (6.0912.110): The measuring cell cable is led through the guide chain of the

815 and the two black plugs are inserted into the openings marked

Cond. Cell

of 712

Conductometer, the two red plugs into those marked

Pt 100/1000

. Later (as soon as the titration

head has been mounted) the measuring cell with the SGJ sleeve (6.1236.050) will be inserted
in the SGJ 14 opening of the titration head.