3 error messages, Error messages – Metrohm 860 KF Thermoprep User Manual
Page 32
4.3 The display
■■■■■■■■■■■■■■■■■■■■■■
24
■■■■■■■■
860 KF Thermoprep
Symbol
Significance
An error has occurred. The type of the error will be iden-
tified in the status display.
4.3.3
Error messages
Table 2
Error messages
Error message
Cause
Remedy
Flow rate error
The tubing system
is leaking or the
flow rate is set too
high.
Screw all of the tubing
connections and the screw
connectors of the drying
flasks tightly.
Check the functioning of
the pump.
In cases of external
gas supply, the
pressure is either
too low or too
high.
Adjust the pressure of the
external gas to 0.1…0.3
bar.
The flow sensor is
defective.
Inform a service techni-
cian.
Max. temp.
reached
The oven tempera-
ture is greater than
265 °C. There is
presumably a
defect in the tem-
perature control.
Switch off the instrument
without delay and allow
the oven to cool down.
Inform a service technician
of the problem if the error
appears again after switch-
ing back on.
Adj. data missing The adjustment
data of the instru-
ment are not cor-
rect.
The instrument must be
readjusted. Inform a serv-
ice technician.
- 915 KF Ti-Touch (382 pages)
- 800 Dosino (53 pages)
- 767 Calibrated Reference (23 pages)
- 940 Professional IC Vario ONE/SeS/Prep 2 (54 pages)
- 754 Dialysis Unit (49 pages)
- 815 Robotic Soliprep for LC (76 pages)
- Vision Manual (207 pages)
- tiamo 2.1 Manual (1532 pages)
- 825 Lab Link (37 pages)
- 808 Titrando (70 pages)
- 902 Titrando (52 pages)
- 756 KF Coulometer (163 pages)
- 756 KF Coulometer (162 pages)
- 940 Professional IC Vario ONE/LPG (98 pages)
- 850 Professional IC Anion MCS Prep 3 (154 pages)
- 850 Professional IC Anion MCS Prep 3 (152 pages)
- 904 Titrando (58 pages)
- 850 Professional IC Anion MSM-HC MCS Prep 2 (150 pages)
- 930 Compact IC Flex Oven/ChS/Deg (47 pages)
- 872 Extension Module Liquid handling (64 pages)
- 814 USB Sample Processor (90 pages)
- 814 USB Sample Processor (91 pages)
- 940 Professional IC Vario (43 pages)
- Vision – Tutorial (40 pages)
- 799 GPT Titrino (242 pages)
- 889 IC Sample Center (68 pages)
- 761 Compact IC (228 pages)
- 851 Titrando (100 pages)
- 748 DH Sample Changer (32 pages)
- 940 Professional IC Vario ONE/SeS/HPG (51 pages)
- 896 Professional Detector – Amperometry (62 pages)
- 877 Titrino plus (139 pages)
- 881 Compact IC pro – Anion (129 pages)
- 940 Professional IC Vario ONE/ChS/HPG (112 pages)
- 930 Compact IC Flex Deg (41 pages)
- 840 PC Control 5.0 / Touch Control (351 pages)
- 940 Professional IC Vario ONE/Prep 1 (45 pages)
- 776 Dosimat (42 pages)
- 717 Sample Changer (36 pages)
- 815 Robotic USB Sample Processor XL (113 pages)
- 815 Robotic USB Sample Processor XL (114 pages)
- 940 Professional IC Vario ONE/SeS/PP (126 pages)
- 838 Advanced Sample Processor Installation Instructions (109 pages)
- 700 Dosino (55 pages)
- 719 S Titrino (152 pages)