beautypg.com

Figure 12, Connecting the heating tubing – Metrohm 860 KF Thermoprep User Manual

Page 23

background image

■■■■■■■■■■■■■■■■■■■■■■

3 Installation

860 KF Thermoprep

■■■■■■■■

15

2 Use the accompanying knurled screw to screw the 6.2060.010

cable holder tightly onto the guide head from below.

3 Guide the heating tubing cable through the eyelet of the cable

holder.

Note

The cable holder prevents the cable of the heating tubing from
coming into contact with parts of the housing that are hot.

4 Connect the heating tubing cable to the Outlet heater connector

on the rear panel of the instrument.

4

Figure 12

Connecting the heating tubing

Rotate the plug in such a way that the three contact pins match the
alignment of the corresponding openings on the socket. Press the
plug against the socket and rotate the front knurled screw in clock-
wise direction.

Note

The heating jacket of the heating tubing is heated up to approx.
40...50 °C as soon as the instrument is switched on. This prevents
the condensation of moisture in the tubing when this is expelled
from the sample and transferred with the aid of a carrier gas into a
KF titration cell.