Troubleshooting, 6 troubleshooting – Metrohm 726 TP Backup User Manual
Page 23
726 TP Backup
22
8.726.8007 Instructions for Use
6 Troubleshooting
Should problems arise during operation of the 726 TP Backup program, please
first check whether
•
the correct cable is used (see section 2.1)
•
the cable is plugged into the correct socket (see section 2.3)
•
the RS232 interface configurations on the computer and the
726 Titroprocessor match exactly.
If the connection is terminated during a transfer, ensure that both the PC and the
726 Titroprocessor can no longer send. In the case of the 726 Titroprocessor,
switch to the dialog page Config. / Interfaces and press – possibly several
times – the softkey [Reset RS232 1] or [Reset RS232 2].
In what follows, several of the most important error messages and warnings are
described.
On starting
It is not possible to start the program twice. Confirm the message with OK.
This message appears if the 726bkup.ini file does not exist. Confirm this
message with OK, the program starts with the default settings. Set the
configuration to suit your requirements (see section 2.3) and save the settings
using the Save command in the Configuration menu.