8 Technical specifications
902 Titrando
■■■■■■■■
35
8 Technical specifications
8.1
Measuring interface
The 902 Titrando has one galvanically isolated measuring interface.
The measuring cycle is 100 ms for all measuring modes.
8.1.1
Potentiometry
One high-ohm measuring input (Ind.) for pH, metal or ion-selective elec-
trodes and one measuring input (Ref.) for separate reference electrodes.
Input resistance
> 1 · 10
12
Ω
Offset current
< 1 · 10
–12
A (under reference conditions)
Measuring mode
pH
Measuring
range
–13 - +20 pH
Resolution
0.001 pH
Measuring
accuracy
±0.003 pH
(±1 digit, without sensor error, under reference conditions)
Measuring mode
U
Measuring
range
–1,200 - +1,200 mV
Resolution
0.1 mV
Measuring
accuracy
±0.2 mV
(±1 digit, without sensor error, under reference conditions)
8.1.2
Temperature
A measuring input (Temp.) for temperature sensors of the Pt1000 or NTC
type with automatic temperature compensation.
R (25 °C) and B value can be configured for NTC sensors.
Measuring range
Pt1000
–150 - +250 °C
NTC
–5 - +250 °C
(R (25 °C) = 30,000
Ω and B (25/50) = 4,100 K)
Resolution
Pt1000
0.1 °C