4 connecting an iconnect, Connecting an iconnect, Figure 15 – Metrohm 902 Titrando User Manual

Page 32: Connecting the iconnect

background image

3.5 Connecting sensors

■■■■■■■■■■■■■■■■■■■■■■

24

■■■■■■■■

902 Titrando

NOTE

Always insert the red plug into the red socket. This is the only way
that shielding against electrical interference can be ensured.

3.5.4

Connecting an iConnect

An external 854 iConnect measuring interface can be connected to the
902 Titrando.

Connect the iConnect as follows:

1 Plug the iConnect plug into the iConnect socket of the 902

Titrando.

Take care to ensure that the marking on the plug matches the mark-
ing on the Titrando as shown in the figure.

Figure 15

Connecting the iConnect

The iConnect is detected automatically and entered as measuring
input into the device properties of the Titrando. If an electrode is
connected to the iConnect that is not yet included in the list of sen-
sors for the control software, then a corresponding message will be
displayed.

The iConnect can be plugged in and unplugged while the Titrando is
switched on.

An electrode with integrated data chip, referred to as iTrode, is connected
to the iConnect.

Connect the electrode as follows:

1 Remove the protective caps on the iConnect and the electrode.