beautypg.com

Metrohm 902 Titrando User Manual

Page 41

background image

■■■■■■■■■■■■■■■■■■■■■■

7 Appendix

902 Titrando

■■■■■■■■

33

Assignment

Pin No.

Function

*

Output 8

6

Output 9

7

Output 10

8

Output 11

13

Output 12

19

Output 13

20

0 volts / GND

14

+5 volts

15

0 volts / GND

25

* Signal activated only for operation with Touch Control.

Table 2

Explanation of the individual functions

Function

Explanation

Start

The current method is started at the time of acti-
vation.

t

pulse

> 100 ms

Stop

The current method is canceled (Stop) at the time
of activation.

t

pulse

> 100 ms

Quit

The current command in the determination run
will be canceled at the time of activation.

t

pulse

> 100 ms

Ready

The instrument is ready to receive a start signal.

Conditioning OK

The line is set when Conditioning with SET titra-
tion and KFT titration is at OK. The line remains
set until the determination is started with
[START].

Determination

The instrument performs a data-generating
determination.

EOD

End of Determination.

Pulse (t

pulse

= 200 ms) after a determination or

after a buffer/standard solution during calibration
using a Sample Processor.

Error

The line is set for error message display.