Chapter 5

Maintenance

This chapter covers cleaning, inspection, troubleshooting, and
specifications for the Bird BA–300/310 Blower Assemblies.

Troubleshooting

For corrections requiring repair or replacement of components see the
appropriate section. Only those functions within the scope of normal
maintenance are listed. This manual cannot list all malfunctions that
may occur, or corrective actions. If a problem is not listed or is not
corrected by the listed actions, notify a qualified service center.

WARNING

Never attempt to connect or disconnect RF equipment from the

transmission line while RF power is being applied.

Leaking RF energy is a potential health hazard.

WARNING

Disconnect the unit from all power sources before servicing.

The unit may be energized from multiple sources.

The potential for electric shock exists.

Problem

Possible Cause

Possible Correction

No air flow from
blowers; indicator
light off

Unplugged power cable

Connect the power cable

Fuse burnout

Replace fuse after checking and correcting
the burnout cause. (See “Fuse” on page 13)

No ac power

Make sure ac power is properly connected
and turned on

No air flow from
blowers; indicator
light on

Fan obstructed by bent grill

Straighten the grill

Fan motors overheated

Clean the grill and fan blades (See
“Cleaning” on page 12)

Fan motors burnt out

Replace the fan motors (See “Fan” on
page 14)

Air flow from
blowers; indicator
light off

Lamp burnt out

Replace the lamp (See “Indicator Light” on
page 13)