Figure 4 thermoswitch assembly, Connecting rf power, Using a “qc” connector coupling – Bird Technologies 8890-320 Series User Manual
Page 19: Using a 13-30 coupling

7
6.
Securely solder the control switch leads to the lugs (G) of the connector base.
Note: The ring-nut (A) must be in place over the base plug (D) with the knurled
end facing out.
7.
Screw on the cover ring, then fasten the cable clamp (E) in place and tighten both
yoke screws (H).
8.
Put the plug back on the thermoswitch and tighten the nut (A).
Figure 4 Thermoswitch Assembly
Connecting RF Power
Note: After installing the load, the RF transmission line can be attached using standard coaxial line cou-
pling kits.
Using a “QC” Connector Coupling
Use a 50 ohm coaxial cable such as RG-218/U or -220/U (-17A or -19A), appropriate for the frequency and power
level of operation.
Note: Use a cable connector which will mate with the one on the load.
Using a 13-30 Coupling
Use 50 ohm coaxial cable such as RG-8A/U, RG-9U, RG-213/U, or equivalent with a male 13-30 plug.
WARNING
Never attempt to connect or disconnect RF equipment from the transmission line while RF power is being
applied. Leaking RF energy is a potential health hazard.