2. Before operating the unit

Precautions when installing the unit

1. Carefully select an installation site.

Take special care not to install the unit at a place described below:

Uneven surfaces or dirty surfaces

Where flammable gas or corrosive gas exists

Where the ambient temperature is 35℃ or more

Where temperature changes severely

Where humidity is high

Where subject to direct sunlight

Where vibration is severe

Install this unit at a place with spaces shown below.

15cm

or

more

15cm

or

more

Front

15cm or more

1m or more

2. Install the unit on a level surface.

Install the unit on a level surface. If the whole bottom surface of the unit does not contact
the surface evenly, vibrations or noises may result. This might cause unexpected troubles
or malfunctions.

The unit weight is approx. 45 kg.
When lifting the unit for transportation and installation, carefully handle it by at least two
people.

3. Installation

The unit might fall down or move by an earthquake or an impact resulting in a personal
injury. We recommend making safety measures such as to avoid installing the unit at a
place other than busy places.