4. Operating procedures

Operating procedures (fixed temperature operation)

How to start fixed
temperature operation

MEASURED TEMP.

HEATER

RUN

SET TEMP.

1.Turn the ELB ON. (Turn the ELB to “ON.”)

When the ELB is turned ON, the initial values will be

displayed for about four seconds, then the initial screen
will appear and the current bath temperature and the
previous set temperature are displayed on each of the
indicators.

Measured temperature screen: Indicates the current bath
temperature
Set temperature screen: Displays the previous set
temperature

STOP

RUN

MEASURED TEMP.

HEATER

RUN

SET TEMP.

TIMER

SUB MENU

2. Setting the temperature

Set a temperature using the ▼▲ keys.

STOP

RUN

MEASURED TEMP.

HEATER

RUN

SET TEMP.

TIMER

SUB MENU

3. Starting operation

Press the RUN/STOP key longer.
Fixed temperature operation will start and the RUN lamp
and the heater lamp come on.

4. Stopping operation

Press the RUN/STOP key longer.
Operation stops, the RUN lamp goes off and the screen
switches to the initial setting screen.

When you want to correct
setting errors or change
settings

When you want to change settings, press the ▼▲ keys on
the current screen to enter the setting mode where you can
change settings. Blinking stops in three seconds after
change and setting is completed.

Caution

When you want to lower the set temperature during fixed value operation, note
that it takes some time to reach the reset temperature since the unit has no
cooling capacity.

Immediately after operation has been stopped, the temperature in the bath is
around the set temperature. Operation stop refers only to machine stop and
time needed for decreasing the temperature in the bath is not considered.