beautypg.com

Operating procedures, Operating procedures (auto stop operation), Procedures for auto stop operation – Yamato Scientific DG400 Drying Ovens for Laboratory Tools User Manual

Page 21: About the timer function

background image

19

4. Operating procedures

Operating procedures (auto stop operation)


This mode automatically stops fixed value operation after a certain time from its start set with the
timer.

Procedures for auto stop
operation

STOP

RUN

MEASURED TEMP.

HEATER

RUN

SET TEMP.

TIMER

SUB MENU

1. Setting a stop time

After confirming the temperature you want is set, press
the TIMER key to display characters AStP

on

the measured temperature screen that indicate auto
stop operation.
The set time is displayed on the set temperature
screen.

Set a time you want using the ▼▲ keys.
Pressing the▼▲ keys makes the set time blink. The
time is determined when blinking stops.

About the timer function

The maximum time that can be set for the timer is 999 hours
50 minutes.
Up to 99 hours 59 minutes, time can be set in minutes.
One hundred hours and over are set only in 10 minutes.
Keep the ▼▲ keys pressed to continuously change set
time and you can quickly reach the time you want. Press the
▼▲

keys once at a time for fine adjustment.

STOP

RUN

MEASURED TEMP.

HEATER

RUN

SET TEMP.

TIMER

SUB MENU

2. Starting timer operation

When the time you want is set, press the RUN/STOP key
for about one second while characters AStP

that

indicate auto stop operation are displayed on the
measured temperature screen and the set time on the set
temperature screen.
The RUN lamp blinks and timer operation is started.
Timer starts counting when the temperature in the bath
reaches the set temperature.
Once timer counting is started, the set temperature screen
changes to the remaining time display.

STOP

RUN

MEASURED TEMP.

HEATER

RUN

SET TEMP.

TIMER

SUB MENU

3. Stopping and ending timer operation

Operation stops automatically when the set temperature
has elapsed.
Characters End

blink on the set temperature

screen to indicate operation has ended.

Press the RUN/STOP key for approx. one second to end
the timer operation mode. The screen switches to the