28

Operation Method

Temperature Output Terminal

Specification

Temperature Output

(ANALOG)

• The voltage (DC) corresponding to the measured temperature is output.
• Output temperature range: CLH300 -5 to 85℃, CLH400/600 -20 to 85℃
• Output voltage: 4 to 20mA DC
• Load:

600Ω or bellow

• Resolution: ±1℃
• Connection: M4 screw terminal block

CLH300

Temperature Output

-15

8.75

32.5

56.25

80

4

8

12

16

20

Output Voltage (mA)

Temp.

CLH400/600

Temperature Output

-20

5

30

55

80

4

8

12

16

20

Output Voltage (mA)

Temp.