PID Sensor Tab • 26

NOTE: Make sure you have downloaded the data from the EAGLE 2

to avoid erasing its data when updating the instrument using
the Transmit New Data button.

2. If no user-defined PID gas has been previously defined for the

connected instrument, the name and long name will both appear as
PID for the High Range and Low Range tabs. The full scale, digit,
and factor values will appear as shown in Table 4 below. The rest
of the fields will contain asterisks (*****).

3. Export the current data by pressing the “Export csv file” button.

Choose the file path you wish to save the file in.

Table 3: Default Values for User Defined PID Gas

Full Scale (ppm)

Digit

Factor

High Range

2000

2

1.00

Low Range

50.00

0.02

1.00

Figure 18: Save As