21 • Conversion Table Tab

Data Loader Program. The first value is the gas number, the second
is the gas name, etc.

6. Edit the values you wish to change and save and close the file.

7. Return to the Maintenance Data Loader Program with the User-

Defined Table tab still up and press “Import csv file”.

8. Choose the file you just edited and press “Open”.

9. The values you entered in the csv file will appear in the User-

Defined Table.

10.If the program finds anything wrong with the values that were

entered, the box containing those values will turn red. Make sure
that you have entered valid characters for each field.

11.Once all of the data you wish to enter has been entered and is

correct, transmit the new data to the connected instrument by
pressing Transmit New Data.