Measures for storage or long-time shutdown, Recommendable spare parts list – RKI Instruments SD-705EC User Manual
Page 28

PT2E-1140
4−10 Measures for Storage or Long-time Shutdown
(1) Store the sensor as attached to the detector in a place not exposed to dust and water
splash.
The warm up time will be longer if the power is not supplied to the sensor for a long time.
If the sensor is kept for more than one month without power supply, gas calibration may
be required. So, it is recommendable to supply the power always.
(2) Storage conditions
Temperature :
-10 ∼ +40℃
Humidity
:
30 ∼ 80%RH 以下
Environmental conditions:
Place without organic solvent and gas generation.
4−11 Recommendable Spare Parts List
No.
Description
Checking
interval
Replacement
interval
Q’ty/Detector
1
Sealing packing
1 year
5∼6 years
1 pce
2
Packing
1 year
5∼6 years
1 pce
3
Sensor
6 months
1 year
1 pce
4
Printer circuit
board(Display)
−
5∼6 years
1 set
5
Printed circuit board(Main)
−
7∼8 years
1 set
6
Fuse
−
8 years
1 pce
*1:The above replacement interval may vary depending on the conditions
of actual use. It does not mean the warranty period.
The replacement interval also may vary by the result of regular inspection.
*2 : The main reason of the printed circuit board is the deterioration of capacitors.
27