4−6 Alarm (Transmission) Test Method
Alarm functions can be confirmed.
When make alarm test (transmission test), announce it to respective
department beforehand. Carry it out after making proper treatment.
!
CAUTION
①
Press SET switch● in menu display .
② Press UP switch ▲ or DOWN switch ▼ to select whether alarm contact is
activated or not.
(No operation)⇔ (Operation)
③ Press SET switch● to decide.
④ When alarm test mode is entered, test level (zero value) is flickered on LCD
display .
When “ON” is selected at this time, ZERO light and SPAN light are flickering
simultaneously.
⑤ To press UP switch▲ or DOWN switch▼, test level(between 0 to full scale, over
scale) can be changed.( 4-20mA output is also changed according to indication).
⑥ Test level exceeds 1
st
alarm point, 1
st
alarm is activated.
(After alarm delay time passed, AL1 light is lighting, and 1
st
alarm contact is activated
if “ON” is chosen).
⑦ Test level exceeds 2
nd
alarm point, 2
nd
alarm is activated.
(After alarm delay time passed, AL2 light is lighting, and 2
nd
alarm contact is activated
if ”ON” is chosen).