Masignal output adjustment method, Alarm point change method – RKI Instruments SD-705EC User Manual
Page 24

PT2E-1140
4−4 4mASignal Output Adjustment Method
① Press SET switch● in MENU display
② When signal output adjustment mode is entered is displayed on LCD,
and signal output becomes 4mA (value : zero).
③ Adjust indication value to 4mA(zero) on indication part* of DSC and etc which is
connected separately by pressing UP switch▲ and DOWN switch▼.
④ After adjustment, press SET switch● to decide. When it is decided, SPAN light is
lighting.
* : If indication value cannot be confirmed on indication part, signal output can be
confirmed by connecting a tester (Ammeter) to check pins between TP1(+) and
TP1(-).
4−5 Alarm Point Change Method
1
st
alarm and 2
nd
alarm can be alternated individually.
① Pres SET switch● in menu display
② When alarm set mode is entered, AL light is on and current 1
st
alarm value is displayed
on LCD.
③ When changing alarm point, press SET switch●。
④ Then, AL1 light is changed from lighting into flickering, and alarm point on LCD is
changing as same. Press UP switch▲ and DOWN switch▼ to adjust.
⑤ After adjustment, press SET switch● and return to ①
⑥ Then, press UP switch▲.
⑦ Current 2nd alarm point is displayed on LCD. And AL2 light is lighting.
⑧ 2
nd
alarm point can be changed as same as 1
st
alarm point.
23