43 • Connecting Instruments to the PC Controller Program
9.
Turn on each EAGLE 2 by pressing and holding the POWER ENTER RESET
button on each EAGLE 2 until you hear a beep, then release it. The EAGLE 2
will begin its power up sequence. If a successful connection between the
EAGLE 2 and the SDM-E2 occurs, the home screen will appear on the EAGLE
2 display at the end of the start up sequence. It indicates the calibration gas
value for each channel. If the charge LED was amber, it will begin to blink
green and the battery charging will stop until the EAGLE 2 is turned off.
10. For each instrument, the SDM Docking Station PC Controller Program will
indicate that an EAGLE 2 was found and it will download the instrument’s
parameters. The BUMP
T
and CAL
S
LEDs will flash amber while the data is
being downloaded. Once the parameter download is complete, the BUMP
T
and CAL
S
LEDs will be solid amber. When all EAGLE 2 instruments have
been turned on, connected, and downloaded, the main program window will
appear as shown below. When the EAGLE 2 is connected to the PC Controller
Program, the date and time of the instrument are automatically updated to the
current date and time on the PC Controller Program’s screen.
NOTE:
Turn each EAGLE 2 on 10-15 seconds after the previous instrument to
minimize the possibility of the EAGLE 2 not connecting to the PC
Controller Program. If an EAGLE 2 does not connect and the pump
continues to run or IrDA CONNECT appears on the EAGLE 2 display,
use the instrument’s POWER ENTER RESET button to turn it off and
turn it back on again.
CH4 50%LEL
OXY 12.0vol%
H2S 25.0ppm
CO 50ppm
---
---