Setting up on the laboratory table

Fig.4

You can also mount the SR 200 on your laboratory table.

Fig.4

* Measure thickness of laboratory table. The thickness

should not be less than 30mm to ensure that safe working

with the SR 200 is not adversely affected.

* Use screws with a max. diam. of 8mm and a length of

laboratory table thickness + 25mm

* Stand the SR 200 on the table
* Draw holes for the table using the holes present in the

motor of the SR 200

The distance from the holes to the front edge of the table

must be such that it is possible without difficulty to mount and

dismount the collecting pan or the filter hose.

Electrical connection

The electrical connection may only be made by an electrician.

*

The voltage and frequency of the SR 200 can be found on the nameplate.

* Ensure that the values agree with the applicable power

supply.

* Connect the SR 200 to the power supply using the

connection cable supplied.

The power cable supplied does not have a plug because the

type of plug depends on the setting-up location and the

relevant regulations in the country concerned
* When connecting the power cable to the power supply,

provide an external fuse in accordance with the regulations

of the setting-up location.

If the values on the nameplate are not observed, electronic and
mechanical components may be damaged.

Before operating for the first time, the direction of rotation must be
checked, see rotation arrow on the motor.
If the direction of rotation is incorrect the grinding will be inadequate
and mechanical parts may be damaged.