Setting up on the laboratory table
Fig.4
You can also mount the SR 200 on your laboratory table.
Fig.4
* Measure thickness of laboratory table. The thickness
should not be less than 30mm to ensure that safe working
with the SR 200 is not adversely affected.
* Use screws with a max. diam. of 8mm and a length of
laboratory table thickness + 25mm
* Stand the SR 200 on the table
* Draw holes for the table using the holes present in the
motor of the SR 200
The distance from the holes to the front edge of the table
must be such that it is possible without difficulty to mount and
dismount the collecting pan or the filter hose.
Electrical connection
The electrical connection may only be made by an electrician.
*
The voltage and frequency of the SR 200 can be found on the nameplate.
* Ensure that the values agree with the applicable power
supply.
* Connect the SR 200 to the power supply using the
connection cable supplied.
The power cable supplied does not have a plug because the
type of plug depends on the setting-up location and the
relevant regulations in the country concerned
* When connecting the power cable to the power supply,
provide an external fuse in accordance with the regulations
of the setting-up location.
If the values on the nameplate are not observed, electronic and
mechanical components may be damaged.
Before operating for the first time, the direction of rotation must be
checked, see rotation arrow on the motor.
If the direction of rotation is incorrect the grinding will be inadequate
and mechanical parts may be damaged.