Damper System Operation
16
Interlocks
Introduction
Interlocks
The test controller incorporates interlock circuitry that can automatically stop a
test when a system problem, specimen failure, or an operator error could result in
unwanted or unexpected actuator movement. The test controller interlock
circuitry will cause one of two interlocks to occur in response to specific
conditions encountered during testing. These interlocks are:
•
Hydraulic Interlock – stops the program and removes hydraulic pressure
from the system.
•
Program Interlock – stops the current test program when certain test
parameters are not met or are exceeded.
The following table lists the possible types of interlock conditions for the test
controller.
Condition
Cause
Type of Interlock
Emergency Stop
Emergency Stop or E-Stop button pressed
Hydraulic
Program Aux
Open circuit at the Prog Intlk rear panel connector
Program
Hydraulic event
Fault in the performance of hydraulic components
Hydraulic
Mechanical event
Fault in the performance of mechanical components
Hydraulic
End-of Count
Completion of a preset number of cycles in the current test
program
Hydraulic or
Program
Underpeak
Selected signal fails to reach a preset minimum or maximum
level
Hydraulic with
indicator option
*
*
These interlocks can be set to indicate only. For further information on configuring these interlocks,
refer to the test controller documentation.
Error
When the control channel's DC error signal exceeds a preset
maximum level
Hydraulic with
indicator option
*
Upper Limit
Lower Limit
Transducer feedback signal exceeds a preset minimum or
maximum level
Hydraulic with
indicator option
*