MTS Axial Extensometers User Manual
Page 49

Calibration
49
3. Adjust compression.
A. Remove the zero pin, fixture, or release the zero stop.
B. Adjust the calibrator between zero and 100% compression of
the extensometer's full-scale range three times. This exercises
the extensometer to remove any hysteresis.
C. Adjust the calibrator for the 80% compression setting (or the
range being calibrated) and note the voltmeter reading.
•
If the meter reading is less than -8 V DC you can adjust gain.
Adjust the gain control to set the extensometer signal to
-8 V DC (or 80% of your calculated output).
•
If you cannot adjust the gain control to set the extensometer
signal to -8 V DC, you can change the excitation voltage.
4. Readjust the zero and compression.
Repeat Step 2 and Step 3 until the zero and gain outputs can be
measured without readjustment.
5. Adjust tension.
A. Adjust the calibrator between zero and 100% tension of the
extensometer’s full-scale range three times. This exercises the
extensometer to remove any hysteresis.
B. Adjust the calibrator for 80% tension (or for the maximum
positive travel) of the range you are calibrating. Note the
meter reading.
•
If the extensometer signal is above +8 V DC (or your
calculated output), adjust the
Δ
K (delta K) control for an
extensometer signal of +8 V DC.
•
If the extensometer signal is below +8 V DC (or your
calculated output), the
Δ
K adjustment cannot be made. Return
the
Δ
K adjustment to its original setting and use the gain
control to increase the voltage halfway from its present value
to +8 V DC (this splits the difference between compression
and tension).