Handset controls quick reference guide, Handset controls quick reference guide 8 – MTS Model 494-05 Handset User Manual
Page 8
![background image](/manuals/568027/8/background.png)
MTS Model 494.05 Handset
Overview
8
Handset Controls Quick Reference Guide
*MC SN1 Ch 2
>Displ 130.5
-Force 234 kN
Strain 0.738 mm/m
*
MC SN1 Ch 2
Display Pages:
MC = Manual Command
AO = Auto Offset
HC = Hydraulic Control
Station Number
(interlock number
assigned to station)
Channel
Name
Page Forward
Page Back
Hydraulics Indicator
ON = high or low hydraulics
OFF = hydraulics off
Interlock Reset/Override
Each time you press Reset/
Override, the interlocks are
overridden for 10 seconds
(10 presses=100 seconds).
Program Stop
(active at all times)
Enter
Press to make an item
marked with the scroll
pointer (–) active.
> = active selection
– = scroll position
Exclusive Control
Press to activate exclusive
handset control of the station.
When the handset has
exclusive control, the Station
Manager application cannot
initiate actuator movement.
By default, the Station
Manager application must be
at the Operator access level
for handset operation. You
can change this so the
handset can operate at any
access level.
Program Run
Runs a previously loaded
program or the function
generator.
Thumbwheel
When Manual Command is
active, use the thumbwheel to
position the actuator.
Manual Command (MC)
Press to turn on thumbwheel
control of the actuator using the
selected active-mode signal (>).
Hydraulics OFF
Press to turn off all HSMs and
HPUs assigned to the station
(active at all times).
Interlock Indicator
ON = interlock condition
OFF = no interlock
Blinking = interlock override
Program Hold
Scroll Down
The scroll pointer (–) indicates the
position as you scroll down a list.
*
= Exclusive Control
Select
Display
Pages