Handset Setup Procedures
MTS Model 494.05 Handset
11
Model 494.05 Handset Hwi File Editor Settings
To initially install the handset:
1. Physically install the serial boards required for the number of handsets used
by the system:
The Model 494.05 Handset requires one of the following serial boards:
•
FlexTest 40, 60, 100, 200 systems—a Model 494.47 Dual UART/
Digital Transducer Conditioner mezzanine card that is installed on a
Model 494.40 I/O Carrier board can support up to two handsets.
•
FlexTest IIM, GT, and TestStar IIM systems—a Model 498.71
GRES III board can support up to four handsets.
2. Use the Hwi File Editor application to add the required serial board
resources to the .hwi file.
A. Click the Controller icon.
B. Click the Board tab and add the GRES III board. GRES III boards are
typically placed in VME Bus slot 10.
or
For Series 494 systems, right-click an empty 494.40 I/O carrier slot and
select the Model 494.47 UART mezzanine card.
3. Define the number of handsets used with the system.
A. Click the Controller icon.
B. Click the Miscellaneous tab and select the # of Handset Controls
from the drop-down list.
4. Click each handset icon to select the card and the serial port used by the
handset.
5. Click the serial board icon and set a 19200 baud rate for each Serial Comm
port assigned to a handset.
6. Set the comm port Type for Model 494.47 UART cards to 4 Wire.