Anchoring the load unit – MTS Landmark Floor Standing User Manual
Page 55

Loosen the Manual Crosshead Locking Bolts
Description
Item
136 N·m (100 lbf·ft)
1
271 N·m (200 lbf·ft)
2
271 N·m (200 lbf·ft)
3
Loosen the Manual Crosshead Locking Bolts
Description
Item
190 N·m (140 lbf·ft)
1
3. Cycle the lock control to unclamp, and then reclamp, the crosshead.
4. Use the lock control to unclamp the crosshead. Wait 30 seconds for the pressure in the crosshead locks
to drop to zero.
5. Use the lift control to position the crosshead where you want it.
6. Use the lock control to clamp the crosshead to its current position. Wait 30 seconds for the locks to fully
clamp the columns.
Anchoring the Load Unit
In some situations, it might be necessary to anchor the load unit to the floor. It is possible, though uncommon,
for the load unit to creep during some types of testing. An example would be high frequency testing within
the floor’s resonant frequency range. Contact MTS for additional information.
MTS Landmark™ Floor Standing Load Units Product Information 55
Installation