14. It makes installation easier to route and secure the safety switch sensor cable prior to mounting the side

panel.

a) Place loop clamps over the sensor cable and position the clamps near the provided threaded mounting

holes.

b) If the safety switch is not pre-mounted, position the safety switch with the cable in a downward direction,

and the arrow-on switch pointing forward. Align the safety switch parallel and centered vertically. Using
an M4 x 25mm sockethead cap screw and M4 washer, mount the switch to side panel and torque to
0.57 N-m (5 lb-in.)

c) Loop the cable back along the bottom right edge. Loosely mount the cable to the bottom cross edge

with M4 x 8mm buttonhead cap screws. Slide the cable in the loop clamps to take up the slack. Torque
to 0.57 N-m (5 lb-in.)

d) Mount the side panel.

e) Prior to mounting the rear panel assembly, place the loop clamps onto the sensor cable at approximate

positions to the threaded mounting holes on back of the baseplate.

f) Loosely mount the cable to the baseplate. Slide the cable in the loop clamps to take up the slack.

Torque to 0.34 N-m (3 lb-in.)

8 MTS Acumen™ Electrodynamic Test System

Test Area Enclosure Installation