a) Loosen the top and bottom M4 flathead screws located inside the door latch vertical edge.

b) Adjust the latch so that the upper and lower catch ball bearings touch the latch simultaneously. You

can view this from the side.

c) Slightly tighten the two screws loosened in step 20a such that the latch is somewhat secure.

d) Close and open the door two or three times to allow the latch to align exactly with the catch ball bearings

on the side panel.

e) Fully tighten the two screws loosened in step 20a.

27. Remove the protective covering from the panels and door.

16 MTS Acumen™ Electrodynamic Test System

Test Area Enclosure Installation