3 power failure – Memmert ICP Compressor-cooled incubators User Manual

Page 34

background image

34

D30374 | Date 10/2014

Malfunctions, warning and error messages

Error description

Cause of errors

Rectifying errors

See

Error message AI E-3 in

the temperature display

TEMP

AI E-3

Set 37.0

°C

37

.4

°C

Temperature monitoring

sensor defective. The oper-

ating sensor takes over the

measurement function.

The appliance can

temporarily be

kept in service

Contact customer

service as soon as

possible

page 2

Error message E-3 in

the temperature display

TEMP

Set 45.0

°C

E-3

°C

Operating and monitoring

sensor defective

Switch off appli-

ance.

Remove the cham-

ber load

Contact customer

service

page 2

When switching on

the appliance, the start

animation is displayed

in another colour than

white

Cyan

:

Not enough storage

space on the SD card

Red

:

The system files could

not be loaded

Orange

: The

fonts and images could

not be loaded

Contact customer

service

Contact customer

service

Download the firm-

ware update at mem-

mert.com and install it

page 2

page 2

6.3 Power failure

In case of a power failure, the appliance operates as follows:

In manual mode

After power supply has been restored, operation is continued with the parameters set. The

time and duration of the power failure are documented in the log memory.

In timer or programme mode

In timer or programme mode

In case of an interruption of the power supply of less than 60 minutes, the current pro-

gramme is continued from the point at which it was interrupted. For longer interruptions of

the power supply, all appliance functions (heating, fan etc.) are switched off.

In remote control mode:

The previous values are restored. If a programme has been initiated via remote control, it is

continued.